Ul adapter factors, Lv models rv-ha-hb models – Brookfield UL Adapter User Manual

Page 4

Advertising
background image

Brookfield Engineering Laboratories Inc.

- 4 -

Operating Instructions M/91-080-I1003

The factor is used to calculate viscosity in units of centipoise (cP) or milli Pascal seconds (mPa•s) when using Dial
Reading or Model DV-I Viscometers. The Viscometer dial reading/display value x Factor = cP (mPa•s).

UL Adapter Factors

Figure 5: Pivot Cup Compatibility

TYPE I

TYPE II

TYPE III

Threaded

Hole

When an older dial reading visco-
meter is used, the viscometer Pivot
Cup may need to be replaced with
the current design. If your visco-
meter has a Type I or Type II pivot
cup, the cup should be replaced
with a Type III cup. Contact
Brookfield or your Brookfield agent
for information.

LV Models

RV-HA-HB Models

Shear

Shear

Factor

Speed

Rate

Speed

Rate

(RPM)

(sec

-1

)

Factor

(RPM)

(sec

-1

)

RV

HA

HB

60.0

73.38

0.1

100.0

122.30

0.64

1.28

5.12

30.0

36.69

0.2

50.0

61.15

1.28

2.56

10.24

12.0

14.68

0.5

20.0

24.46

3.20

6.40

25.60

6.0

7.34

1.0

10.0

12.23

6.40

12.80

3.0

3.67

2.0

5.0

6.12

12.80

25.60

1.5

1.83

4.0

4.0

4.89

16.00

0.6

0.73

10.0

2.5

3.06

25.60

0.3

0.37

20.0

2.0

2.45

1.0

1.22

0.5

0.61

Shear Rate = RPM x 1.223