Brookfield Engineering Laboratories Inc.

-  -

Operating Instructions M09-086

Assembly & Operation

1. Attach the mounting channel to the viscometer by threading the upper mounting screw (HT-19) into the

viscometer pivot cup. Do not overtighten. The end of the mounting channel with the pin must be at the

bottom, as shown in Figure 1.

2. Open Tube Operation (#304 stainless steel)
The water jacket (ULA-49EAY) must be removed when the “open tube” method is used. This allows

measurement in a beaker or other container. The sample chamber must be inserted into the clamping collar

before immersion in sample fluid. The top of the sample chamber should be flush with the top edge of the

clamping collar.

The clamping collar connects to the mounting channel as shown in Figure 1. Attach the spindle to the

viscometer by raising the spring-loaded sleeve at the base of the viscometer pivot cup and inserting the

spindle so that the bottom of the coupling is flush with the bottom of the sleeve. Slide the sleeve back down

to hold the spindle in place. Slowly lower spindle/chamber assembly into sample fluid. Observe and ensure

that the fluid level meets the immersion groove located on the outside of the sample chamber as shown in

Figure 2.

3. Closed Tube Operation with Water Jacket
The UL Adapter may be used with the ULA-49EAY water jacket and a circulating water bath by connecting

the water jacket to the bath inlet and outlet ports (Figure 3). The tube end cap (ULA-34) is snapped over the

bottom end of the sample chamber before the sample is added. Be sure that it is securely seated in the groove.

Fill with 16 mL of sample fluid. Insert the clamping collar into the mounting collar as shown in Figure 1.

Insert the sample chamber upwards into the water jacket and clamping collar. When fully inserted, secure the

sample chamber by tightening the clamping knob. Attach the spindle to the viscometer by raising the spring-

loaded sleeve at the base of the viscometer pivot cup and inserting the spindle so that the bottom of th coupling

is flush with the bottom of the sleeve. Slide the spindle into the chamber. Attach the Clamping Collar to the

Mounting Channel, slide upwards, and secure in place by tightening the mounting screw. Confirm that the

fluid level in the sample chamber covers the conical portion of the spindle shaft.

CAUTION: The spindle, sample chamber, and cap (if used) should be clean before use. The cap material

is low density polyethylene. Replace when damaged or loose.

CLAMPING COLLAR

CLAMPING KNOB

IMMERSION GROOVE

END CAP GROOVE

Figure 2: Sample Tube Immersion

Chamber Shown is #304 Stainless Steel