Viscometer Operation

Attach the Vane Spindle to your Viscometer as shown in the illustration on page one (Note: spindles

have a left-hand thread for installation). Immerse the Vane Spindle into the fluid past the top of the

Vane to the primary immersion mark on the shaft. Operate your Viscometer per the instructions in the

Operator’s Manual that came with your instrument. Record the instrument Speed (in RPM) and the

Torque Reading (as a percent). The Viscosity is calculated manually using the following equation:

Spindle

Torque

Shear Stress Range

Viscosity Range cP

Range

Pa

dyne/cm

(mPa•s) @ 10 rpm

2

Table 2 - Spindle Range Data

V-71

V-72

LV

.188-1.88

1.88-18.8

104.04 -

1040

V-73

LV

.938-9.38

9.38-93.8

502 -

502000

V-74

LV

9.38-93.8

93.8-938

5.09 -

50900

V-75

LV

3.75-37.5

37.5-375

1996 -

19960

V-71

RV

.5-5

5-50

262 -

2620

V-72

RV

2-20

20-200

1110 -

11100

V-73

RV

10-100

100-1000

5350 -

53500

V-74

RV

100-1000

1000-10000

54300 -

543000

V-75

RV

40-400

400-4000

21300 -

213000

V-71

HA

1-10

10-100

524 -

5240

V-72

HA

4-40

40-400

2220 -

22200

V-73

HA

20-200

200-2000

10700 -

107000

V-74

HA

200-2000

2000-20000

108600 -

1086000

V-75

HA

80-800

800-8000

42600 -

426000

V-71

HB

4-40

40-400

2096 -

20960

V-72

HB

16-160

160-1600

8880 -

88800

V-73

HB

80-800

800-8000

42800 -

428000

V-74

HB

800-8000

8000-80000

434400 -

4344000

V-75

HB

320-3200

3200-32000

170400 -

1704000

V-71

5XHB

20-200

200-2000

10480 -

104800

V-72

5XHB

80-800

800-8000

44400 -

444000

V-73

5XHB

400-4000

4000-40000

214000 -

2140000

V-74

5XHB

4000-40000

40000-400000

2172000 - 21720000

V-75

5XHB

1600-16000

16000-160000

852000 -

8520000

NOT RECOMMENDED FOR USE ON LV TORQUE

Notes: 1) 1 Pa = 10 dyne/cm

2

2) 1cP = 1 mPa•s 3) Possibility of turbulence at speeds above 10 rpm may give artificially

higher viscosity readings 4) 5xHB torque available with DV-III Ultra Rheometer only.

Viscosity (cP) = 100 * TK * SMC * Torque

RPM

where

RPM

= Viscometer spindle speed in RPM

TK

= Viscometer torque constant from Table 3

SMC

= Spindle multiplier constant from Table 4

Torque = Viscometer torque (%) expressed as a

number between 0 and 100

Table 3 - Viscometer Torque Constants

Model

TK

LV

0.09373

RV

1

HA

2

HB

8

5xHB

40

Viscosity (cP) = 100 * TK * SMC * 2xTorque

RPM

Spindle Range Data

If the secondary immersion mark (midpoint of the vane) is

chosen due to small container size, multiply the torque reading

by a factor of two (2).

Table 4 - Spindle Multiplier Constants and Accuracy

Spindle

SMC

Accuracy*

V-71

2.62

±2%

V-72

11.1

±2%

V-73

53.5

±2%

V-74

543

±5%

V-75

213

±3%

* Allowable error percentage is calculated on the

full scale range viscosity for each spindle at a

defined speed.

Advertising