8 Troubleshooting

40

R II Operation Manual, Version E

Table 8-1: General malfunction and their remedy

Malfunction

Possible cause

Corrective measure

Seal is defective

Replace seal

Table 8-2: Malfunctions with vacuum controller and vacuum pump and their remedy

Malfunction

Possible cause

Corrective measure

Frequent switching of valve or pump

System is leaky

Check all sealing points (tubes and
their connections)

Chosen hysteresis is too small

Choose larger hysteresis (if end-
vacuum is higher than 700 mbar,
switch to automatic hysteresis)

Vacuum is not reached

Back evaporation at Rotavapor

Empty receiving flask

Water pressure of water jet pump is
too low

Open water tap completely

Distillation “died out”

Manually decrease the pressure until
the distillation starts again

Distillation stopped, although not
dried out completely

Back evaporation from the receiving
flask is too strong (especially for
solvent mixtures)

Empty receiving flask and restart
distillation

Malfunction in distillation procedure
which is not exactly defined (e.g.
sudden cooling, heat flow too low,
etc.)

Decrease the pressure manually until
the distillation starts again

NOTE
At the R II no errors are displayed and read out via the RS485 interface.

Reset of the overtemperature sensor at the bottom of the Heating bath B-491

Let the bath cool down and empty it. Then
push the temperature sensor reset button at
the bottom side of the heating bath e.g. by
means of a tooth stick.

Fig. 8.1: Heating bath B-491 bottom view