BUCHI Rotavapo R-250 User Manual

Page 17

Advertising
background image

17

Rotavapor R-250

4 Putting into Operation

4.10 Operating the Shut-off tap

The shut-off tap is of a special design. It does not have a
con ti nuous thread on its inside for tightening it, but rather a
sli ding plane with two fixed latching positions. The closing
pres sure when it is in a closed position is provided by a
prestressed spring.

s

ching piece and turn the white lower section of the grip
ቢ clock wise until the tap is tightly seated.

s

Turn the black upper section of the grip

ባ clockwise

until the shut-off tap latches into the first position. If the
opening is not large enough, continue turning until the
tap reaches the second latching position.

4.11 Operating the Inlet Valve

The inlet valve is also closed by an integrated spring, which
applies contact pressure between the PTFE and the ground-
glass joint. The valve can be opened and closed by operating
the lever manually. The setting screw on the lever permits
precise continuous metering or venting.

Fig. 4.14: Shut-off tap

Fig. 4.15: Inlet tap

Advertising