8 Troubleshooting

49

B-390

Operation Manual, Version C

8

Troubleshooting

8 .1

Malfunctions and their remedy

The table below lists possible operating errors and their cause. As remedy set the parameter stepwise
in the opposite direction or fix the missing part.

Table 8-1: Possible cause

Problem

Possible cause

Unstable liquid stream

The liquid flow rate is too low.
The nozzle is not adequately cleaned (frequent cause).
The frequency is too high.
The amplitude is too high.

Unstable bead chain

The frequency is too high or too low.

The liquid flow rate is too high or too low.
The Nozzle is not adequately cleaned.

The amplitude is too low or too high.

Non-homogenous bead-size-distribution

The liquid flow rate is too high.
The frequency is too high.
The electrostatic tension is too low.

The immobilization mixture is a non-Newtonian liquid, making it
difficult to extrude or to prill.

The bead chain does not separate

The electrode is not connected to the control unit.
The electrical tension is too low.
The electrode is not on.

Beads are not visible in the strobo light

The vibration unit is not activated.
The vibration unit is not put on the bead producing unit.
The vibration frequency is too low or too high.
The viscosity of the immobilization mixture is too high.