7 Troubleshooting

37

C-640 Operation Manual, Version A

7

Troubleshooting

Problem

Caused by / Conditions

Solution

After switching the unit on no LED on
the front panel is lit.

Failure of the power line fuse.

Check the fuse.

(The fuse is integrated in the main
power connector on the backside of
the instrument.)

Electronics failure.

Call service center.

Increased noise and drift.

Leaking cell, dirty cell, bubbles,
non-transparent mobile phase, old or
bad lamp.

Perform Cell test with lamp test
included.
Check cell connections.
Check mobile phase.
Check cell purity manually.
Install pressure regulator.

Electronics / Optics failure.

Call service center.

Leaking cell.

Broken window, or leaking seal.

Repair or replace the cell - see
chapter 6.3 “Cleaning and replace-
ment of cell windows”.

Warning or error message.

-

See next two sub-chapters. Try
resetting the error. If resetting does
not solve the problem, call the
service center.

After switching the unit on the
display is not working and the red
LED (ERROR) blinks.

Display electronics / power supply
failure.

Call service center.

7 .1

Internal display error codes

Any failure in conjunction with the internal communication of the detector or its display electronics is
considered as internal error. Any time an internal error occurs the red ERROR LED blinks.

Code

Meaning

Repair

01h

Power is down!

Call service center.

02h

Non-volatile memory failure!

Call service center.

04h

Digital-analog converter failure!

Call service center.

08h

Display failure!

Call service center.

10h

Detector communication failure!

Call service center.

20h

Detector communication timeout!

Call service center.

40h

Interface device failure!

Call service center.