5 – 32

5 Installation

KjelDigester K-446/K-449

Operation Manual

b

Tasks

WARNING

Risk of acid or toxic fume inhalation with a wrong setting of the Scrub-
ber control.
Serious injuries or death can result.
1. Make sure the Scrubber is removing fumes during the complete digestion

process.

1. Press a [STOP] until the Scrubber control menu appears.

2. Adjust the interval time for switching on the Scrubber by pressing b [+/-].

Range 0 – 255 minutes.

3. Press c [ENTER] to confirm the new value.
4. Adjust the interval time for switching off the Scrubber by pressing b [+/-].

Range 0 – 255 minutes.

5. Press a [STOP].

∙The default menu appears.

5.6 User protection shield, installation

NOTE
Install the user protection shield when working with H

2

O

2

or larger amounts of water (>10 mL).

1. Install the user protection shield a on the rack handles b.

∙11057889 User protection shield