3 Technical data

15

K-355 Operation Manual, Version G

3 .2

Technical data

Table 3-4: Technical data K-355

Distillation Unit K-355

Power consumption

max. 2.2 kW

Connection voltage

230 V ± 10% / 200 V ± 10%

Frequency

50/60 Hz

Current consumption (230 V)

8.5 A

Mains connection

3-pole (P, N, E) via power cord

Recovery rate

≥ 99.5%

Reproducibility (RSD)

≤ 1%

Detection limit

≥ 0.1 mg Nitrogen

Environmental conditions

Temperature
Altitude
Humidity

for indoor use only
5 - 35 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
31 °C decreasing linearly to 67% relative humidity at
35 °C

Installation category

II

Pollution degree

2

Dimensions (W x H x D)

400 x 660 x 360 mm

Weight

17 kg

3 .3

Materials used

Table 3-5: Materials used for the K-355

Component

Material designation

Material code

Housing

Polyurethane

PUR / UL V0

Glass parts

Borosilicate glass 3.3

DIN/ISO 3585

Steam generator isolation

Ceramic fiber

Multitherm 550

Steam generator housing

Stainless steel

1.4301

Protective door

Polymethyl methacrylate

PMMA

Condenser cover

Polymethyl methacrylate

PMMA

Connecting stopper

Hypalon

CSM

NaOH tube

Ethylene Propylene Dimonomer

EPDM