Cannon Instrument CMRV-5000 User Manual

Page 103

Advertising
background image

97

CANNON

®

Mini-Rotary Viscometer

CMRV-5000 Instruction & Operation Manual

Version 1.0b—August, 2011; CANNON

®

Instrument Company

2139 High Tech Road • State College, PA • 16803 • USA

Choosing an analysis
4. Select the desired Data Export analysis option. The CMRV Ports

Output Configuration window will open.

5. Click the Define and Open New Configuration radio button

to

define a new analysis configuration.

NOTE

If you have already configured and saved an analysis, its name will
appear in the list box on the right side of the window. If you click on an
existing configuration and click
OK, the analysis will be performed using
the selected configuration settings. It will not be necessary to complete
the remaining steps in this procedure.

6. Click OK. The ASCII Ports Output Configuration window will

appear.

The ASCII Ports Output Configuration window consists of three
tabbed pages:

Date Filter—allows you to select date/time parameters for the
analysis (see Chapter 13 for additional details).

Sample Filter/Error Page—allows you to select which tempera-
ture profiles/errors/technicians/instruments will be included in the
analysis (see Chapter 13 for additional details).

Port Output Format—allows you to select the output port(s), what
sample data will appear in the output analyses, and how the data will
be displayed/sent.

7. Click on the tab corresponding to the filter you wish to set and

complete configuration options per the following instructions:

Advertising