– 6 –

2.2 LCD

Display

Note: The default setting for the four digit displays shows compression

loads as a positive force and tension loads as a negative force. to
reverse these settings, showing compression loads as negative and
tension loads as positive, see section 4.1 for setting f01.

2.3 Display

Indicators

2.4

MAX/MIN display

– 31 –

Overload Output

When the overload condition is triggered, the corresponding overload output
turns on. this can be used to stop a motorized test stand, or an alarm to pre-
vent damage from accidental overload.

When compression overload occurs photo-couple 1 (PC!) turns on, and
allows current to flow between pins 8 and 10.

When tension overload occurs photo-couple 2 (PC2) turns on and allows
current to flow between pins 9 and 10.

If no overload condition exists, PC1 and PC2 should be closed and will not
allow current flow.

Overload occurs at about 120% of the gauge’s rated capacity. this includes
any weight zeroed during any tare operation.

Comparator Output

When the LO limit is reached, photo-couple 1 (PC1) turns on and allows
current to flow between pins 8 and 10.

When the HI limit is reached, photo-couple 2 (PC2) turns on and allows
current to flow between pins 9 and 10.

Refer to section 7 regarding the activation and setting of Comparator
limits.