Model 5i Digital Force / Torque Indicator

User’s Guide


26


11.1 Calibration Password

Select Calibration from the sub-menu. The display will appear as follows:


To set the password, select Enabled , then Set Password. Use the UP and DOWN keys to increment
and decrement the value, from 0 to 9999. When the desired value has been selected, press ENTER, then
ESC to exit the sub-menu.

11.2 Menu Key Password

If enabled, every time the MENU key is selected, a password must be provided. Select Menu Key from
the sub-menu. Follow the same procedure as described in section 10.1.

11.3 Locking Out Other Keys

Other keys may be locked out individually. Select any combination of keys (UNITS, MODE, ZERO,
DATA) by pressing ENTER in the Passwords sub-menu. Pressing a locked key will prompt the message
“KEY PROTECTED” and then revert to the previous screen.

11.4 Password Prompts

If passwords have been enabled, the following will be displayed when pressing the MENU key or
accessing the Calibration section:


Use the UP and DOWN keys to select the correct password, then press ENTER to continue.

If the incorrect password has been entered, the display will appear as follows:

ENTER PASSWORD

(0000 – 9999)

5000

INCORRECT PASSWORD

Reset password

Request code:

XXXX

Press ENTER or ESC

CALIBRATION PASSWORD


*

Disabled

Enabled
Set Password
(0000 – 9999)

5000

Advertising