Checkline MX-3 User Manual

Page 17

Advertising
background image

MX-3 Ultrasonic Thickness Gauge

13

At this point, the MX-3 has successfully calculated it's internal error

factor, and will compensate for this value in any subsequent

measurements. When performing a "probe-zero", the MX-3 will always use

the sound-velocity value of the built-in probe-disc, even if some other

velocity value has been entered for making actual measurements. Though

the MX-3 will remember the last "probe-zero" performed, it is generally a

good idea to perform a "probe-zero" whenever the gauge is turned on, as

well as any time a different transducer is used. This will ensure that the

instrument is always correctly zeroed.

Calibration

In order for the MX-3 to make accurate measurements, it must be set to

the correct sound-velocity for the material being measured. Different types

of material have different inherent sound-velocities. For example, the

velocity of sound through steel is about 0.233 inches-per-microsecond,

versus that of aluminum, which is about 0.248 inches-per-microsecond. If

the gauge is not set to the correct sound-velocity, all of the measurements

the gauge makes will be erroneous by some fixed percentage. The one-

point calibration is the simplest and most commonly used calibration

procedure - optimizing linearity over large ranges. The two- point

calibration allows for greater accuracy over small ranges by calculating the

probe zero and velocity. The MX-3 provides three simple methods for

setting the sound-velocity, described in the following pages.

Advertising