–15 –

7.2 Recalling Stored Values in Mode C (Continuous)

1. Turn on the tension meter and press the RECALL key, You can exit

recall at any time by pressing the POWER key.

2. The display blinks showing the average value (AVG)

of the first measuring period (POS: 1) and the
symbol.

3. Press the RECALL key. The DISPLAY blinks, showing

the last value (LAST) of the measuring period.

4. Press the RECALL key. The DISPLAY blinks, showing

the maximum value (MAX) or the measuring period
and the symbol.

5. Press the RECALL key. The DISPLAY blinks, showing

the minimum value (MAX) of the measuring period
and the

symbol.

6. Press the RECALL key. The DISPLAY blinks, showing the

maximum peak value (PEAKMAX) of the measuring period
and the

symbol.

7. Press the RECALL key. The DISPLAY blinks, showing the

minimum peak value (PEAKMIN ) of the measuring period,
the PEAK indicator and the

symbol.

8. Press the RECALL key. The measured values no. 1 — n

of the first measuring period can be recalled.

9. Press the RECALL key. The next measuring period (POS: 2) is shown on the

DISPLAY, starting with the average value (AVG).

22

D

cN

23

D

cN

24

D

cN

21

D

cN

26

Peak cN

19

Peak cN

Memory Mode CONTINUOUS

Pos:

1 Pos:

2 Pos:

3 Pos:

4

AVG:

22.0 AVG:

12.0 AVG:

12.0 AVG:

12.0

Last:

23.0 Last:

22.1 Last:

22.1 Last:

22.1

MAX:

24.0 MAX:

22.1 MAX:

22.1 MAX:

22.1

MIN:

21.0 MIN:

5.4 MIN:

5.4 MIN:

5.4

PEAK

MAX

:

26.0 PEAK

MAX

:

28.1 PEAK

MAX

:

28.1 PEAK

MAX

:

28.1

PEAK

MIN

:

19.0 PEAK

MIN

:

1.8 PEAK

MIN

:

1.8 PEAK

MIN

:

1.8

10.8

14.2

14.1

15.4

10.0

19.4

11.2

18.3

7.3

22.9

8.9

17.5

6.1

17.3

10.2

7.8

n

n

n

n

– 22 –

o

3rd calibration point:

o

1. Repeat steps 1 through 4 from the preceding procedure using a weight which

corresponds to 70% of the tension range

2. Press the RECALL key. As long as the RECALL key is

depressed, the DISPLAY shows a decimal value which is
higher than the second decimal value, e.g., 8000. This decimal
value may vary from instrument to instrument. Write down the
value.

3. Release the RECALL key when the value shown on the

DISPLAY is stable (the reading might fluctuate greatly).
The DISPLAY shows E 100.

o

4th calibration point:

o

1. Repeat steps 1 through 4 from the preceding procedure using a weight which

corresponds to 100% of the tension range

2. Press the RECALL key. As long as the RECALL key is

depressed, the DISPLAY shows a decimal value which is
higher than the third decimal value, e.g., 9500. This decimal
value may vary from instrument to instrument. Write down the
value

3. Release the RECALL key when the value shown on the DISPLAY is stable (the

reading might fluctuate greatly).

4. Press and then release the RECALL key. The DISPLAY switches off after

approximately 1 second. The new calibration is stored.

5. Press the POWER key. the instrument switches off.

6. Verify the new calibration, following the directions in Sec. 8.0 If this procedure

shows a deviation, you can recalibrate the tension meter again or restore the factory
calibration as described in Sec. 9.3.


If the verification of the calibration according Sec.8.0 shows a deviation beyond
the allowable tolerance and a reliable operation is no longer allowed, the
instrument has to be returned to Electromatic for factory recalibration.