Series TT01 Digital Cap Torque Testers

User’s Guide


12


10.2 Settings
The display appears as follows:


Thresholds and percentage drops are illustrated below:

Function Description

Enabled

If enabled, 2PK appears as one of the operating modes. In the main display, the
Peak readings will reference the first and second peaks – first peak on top,
second peak below. Refer to the Home Screen & Control section for details.

Peak Settings

Press ENTER to access the Peak Settings sub-menu. See the following sections
for details.

Auto Output

Press ENTER to access the Auto Output Settings sub-menu. See the following
sections for details.

Auto Store PK1

Automatically stores the first peak reading to memory.

Auto Store PK2

Automatically stores the second peak reading to memory.

Auto Zero

Automatically zeroes the display following data transmission and/or storage.

Threshold 1

Sets the percentage of full scale at which the first/second peak detection
feature becomes active. This threshold is provided to ignore peaks that can
occur during sample loading and unloading. Available settings: 1-90%, in
1% increments between 1-5%, and in 5% increments between 5-90%.

Percentage Drop 1 Identifies the first peak via detection of a specified percentage drop from

peak.

Threshold 2

Same as with Threshold 1, but refers to a percentage of full scale beyond
the first peak. For example, for a 50 lbFin capacity tester, if the first peak is
20 lbFin, and Threshold 2 is set to 15%, the threshold is equal to 27.5 lbFin.

Percentage Drop 2 Same as Percentage Drop 1, for the second peak.
Auto Zero Delay

Sets the time delay before the primary and peak readings are zeroed.
Available settings: 1-60 sec., in 1 sec. increments between 1-5 sec., and in
5 sec. increments between 5-60 sec.

PEAK SETTINGS

Thresh. 1:

5 %

% Drop 1: 10 %
Thresh. 2: 5 %
% Drop 2: 10 %
Auto Zero Delay

3 sec.