Checkline RH6 User Manual

Page 10

Advertising
background image

Electromatic Equipment Co., Inc.

-10-

5. Now put the third gasket ring into the bottom part.

6. Fit the metal ring on the third gasket.

7. Take the upper part with the RH6 and attach these carefully at the bottom part of

the calibration device.


8. Pick up the RH6 together with the calibration device STRAIGHT and DON´T

TURN IT AROUND. Screw it up like shown in the picture.








Then put the RH6 with the calibration device down on a table carefully and proceed
as follows:

Conditioning the sensor
To achieve best results, let the sensor adjust for two hours. The temperature has
to be between 20°C and 26°C.

If the shown measuring value differs more than the factory tolerance (1.5% r.h.), we
recommend to carry out a recalibration as follows.

Electromatic Equipment Co., Inc.

-11-

11.0 OFFSET ADJUSTMENT


1. Leave the measuring device in the calibration equipment, and switch it on.

2. Press the Rolling Menu button until you reach the main menu.

3. Select the menu item Options by pressing the button and confirm by pressing

.

4. Navigate to

Setting

using the button and confirm by pressing

again.

5. Enter the superuser password using the buttons

resp.

and confirm by

pressing .

The superuser password after consignment is the serial number of the

device, shown on the display after switching on the device or in menu item

Status

.

6. A query if a setting is desired appears. Confirm by

pressing .

7. Wait until the bar has risen completely. The device

adjusts by itself and automatically jumps back to the
measuring window. The adjustment is completed
now.

8. Check the result before you remove the device from

the calibration equipment. Depending on the
temperature the display should show a humidity
around 50% now.


If you made a mistake during the setting, you can reset to the factory calibration as
follows: