INPUT POWER WIRING

WARNING: D

O

NOT

PLUG

THE

POWER

CONNECTOR

INTO

THE

HEADER

WITH

LIVE

POWER

. T

O

DO

SO

WILL

RESULT

IN

PROPERTY

DAMAGE

AND

/

OR

PERSONAL

INJURY

.

WARNING:

IF

A

LITHIUM

BATTERY

IS

REPLACED

WITH

AN

INCORRECT

TYPE

IT

MAY

CAUSE

AN

EXPLOSION

WHICH

WILL

CAUSE

PROPERTY

DAMAGE

OR

PERSONAL

INJURY

.

AC Input Power Wiring (-AC)

WARNING:

DO

NOT

OPERATE

WITH

INCORRECT

LINE

VOLT

-

AGE

.

TO

DO

SO

WILL

RESULT

IN

PROPERTY

DAMAGE

AND

/

OR

PERSONAL

INJURY

. M

AKE

SURE

THAT

THE

POWER

SOURCE

DOES

NOT

EXCEED

240 VAC.

WARNING: I

F

AN

AUTOMATIC

DISCONNECT

DEVICE

IS

USED

ON

THE

AC

INPUT

WIRES

,

THE

DISCONNECT

MUST

ACT

ON

BOTH

THE

LINE

AND

NEUTRAL

WIRES

IN

A

DOUBLE

POLE

,

DOU

-

BLE

THROW

ARRANGEMENT

I

.

E

. DPDT R

ELAY

. U

SING

OTHER

AUTOMATIC

DISCONNECT

ARRANGEMENTS

MAY

CAUSE

PER

-

SONAL

INJURY

AND

/

OR

PROPERTY

DAMAGE

.

The AC power should be supplied by a “clean” primary line,
directly from the power panel. This line should not supply any
other equipment, including the feeding unit, and should be
supplied with a minimum 10 amp breaker.

AC Power Input (See Fig. 7)

Neu (Low)
Line (HI)
Earth Ground

Step 1.

The HI 4050 is configured with a universal power supply rated
from 110 to 240 VAC.

Step 2.

Make sure the VAC power is shut off before installing the
wires to the connector.

Step 3.

Install a 3-wire, minimum 14 AWG power line to the 3-pin ter-
minal block connector. Make sure that the shield is connected
to the local Earth Ground connection.

DC Power Input (-DC)

WARNING:

DO

NOT

OPERATE

WITH

INCORRECT

LINE

VOLT

-

AGE

.

TO

DO

SO

WILL

RESULT

IN

PROPERTY

DAMAGE

AND

/

OR

PERSONAL

INJURY

. M

AKE

SURE

THAT

THE

POWER

SOURCE

DOES

NOT

EXCEED

24 VDC.

CAUTION: Y

OU

MUST

USE

A

POWER

LIMITED

DC

POWER

SUPPLY

(C

LASS

2)

ON

THE

DC

INPUT

WIRING

.

The DC power should be supplied by a “clean” primary line,
directly from the DC power source.

Step 1.

Make sure the VDC power is shut off before installing the
wires to the connector.

Step 2.

Connect the 24 VDC Voltage wire, Ground wire and Shield
wire to the connector that plugs into the DC voltage header at
the rear panel. Make sure that the shield is connected to the
local Earth Ground connection.

Step 3.

Plug the connector into the header at the rear panel. (See
Fig. 7)

STARTING THE HI 4050

Step 1.

Connect the power connector (AC or DC) to the HI 4050.

Step 2.

The Instrument boots up to the Summary Display. (See Fig. 9)

FIG. 9 SUMMARY DISPLAY

Step 3.

Press the Enter button to go to the Configuration menus. (See
Fig. 10)

FIG. 10 CONFIGURATION DISPLAY

Step 4.

Use the HI 4050 Weight Controller User Guide for Configura-
tion, Operation and Troubleshooting instructions.

9440 Carroll Park Drive, Suite 150

San Diego, CA 92121

Telephone: 1-800-821-5831 FAX: (858) 278-6700

Web Address: http://www.hardyinstruments.com

Hardy Instruments Document Number: 0596-0304-01 Rev. C

Copyright September 2006, Dynamic Instruments, All Rights Reserved.