Example – Hardy HI 2204LT Weight Transmitter User Manual

Page 13

Advertising
background image

HI 2201LT/HI 2204LT

9

Step 2.

Install a milli ammeter, in series, into the
output/power loop of the transmitter. If a
voltmeter is to be used for measurements,
and a resistor for calibration is not already
installed in the system, install a 100 ohm,
+- 0.1%, 1/4 W resistor, in series, into the
output/power loop of the transmitter. Volt-
meter readings are show in parentheses.

Step 3.

If a voltmeter is used, attach the voltmeter
across the 100 ohm resistor.

Step 4.

Adjust the transmitter FINE control for a 4
ma (1/4 V) reading on the meter. If neces-
sary, center the FINE control, and adjust
COARSE control first.

Step 5.

Place a know weight on the scale. If possi-
ble, use a full scale weight for best results.

Step 6.

Adjust the SPAN for 20 ma (2V) for full-
scale.

Divide: Calibration Weight/Full-
Scale
Weight = (F)raction of full scale.

Multiply: (F) (16 ma) = 1, the cur-
rent caused by the calibration
weight.

Add: 1 + 4 ma = Calibration Cur-
rent, 1c

Adjust the SPAN to produce the
Calibration Current, 1c, in the out-
put/power loop.

Example

Calibration Weight = 750 pounds
Full Scale Capacity = 1000 pounds

Adjust SPAN for:

(750/1000) (16 ma) + 4 ma = 1c

(0.75) (16 ma) + 4 ma = 1c

12 ma + 4 ma = 1c

16 ma = 1c