4. LOAD CELL CABLE LEGEND

The following legend describes the load cell cable color code with the description of
each colored wire in the cable. Please note that it is important to make sure that the
scale platform is electrically connected to earth ground either through the load cell
cable shield (via the lug terminated green wire) or by directly connecting the scale
platform to earth ground with an additional wire. This will reduce static and other
noise from being induced into the indicator.

For a 6-wire indicator, color code is as follows:

Wire Color

Function

Red

+ Signal

White

- Signal

Green

+ Excitation

Black

- Excitation

Blue

+ Sense

Brown(or Yellow)

- Sense

Green with star lug,(shield)

scale platform ground

For a 4-wire indicator, color code is as follows:

Wire Color

Function

Red

+ Signal

White

- Signal

Green and Blue*

+ Excitation

Black and Brown(or Yellow)*

- Excitation

Green with star lug(shield)

scale platform ground

Note : Connect the +sense and +excitation together as well as the -sense and

-excitation together.