Number of averaging measurements – KEYENCE LS-7600 User Manual

Page 93

Advertising
background image

5-31

Chapter 5 Function Settings

5

Obtaining the object width from Two
Measuring Heads

Set AREA1 for B-EDGE in measuring head
1 and AREA2 for B-EDGE in measuring
head 2.

A1

A2

–1

–1

Setting Method

1 Press the [Up or Down] key and move the cursor to CALC.

Number of Averaging Measurements

Take the following procedure to determine the number of times to average measured
values.

Setting Range

Select the number of averaging measurements from the following.
1, 2, 4, 8, 16, 32, 64, 128, 256, 512, 1024, 2048, and 4096
If you want to average the fluctuation of measured values and obtain stable results,
select a comparatively large number. For high-speed measurement, the number se-
lected should be comparatively small.

2 Press the [ENT] key.

The cursor moves to the coefficient for
AREA1 (A1).

3 Press the [Up or Down] key and set

the coefficient for AREA1 (A1).

4 Press the [Right] key and move the

cursor to the coefficient for AREA2
(A2).

OUT1

CALC

+1

A1 +0A2

AVE

512

MEA-MODE NORMAL
OFFSET

0.00000

OUT1

CALC

+1

A1

+0

A2

AVE

512

MEA-MODE NORMAL
OFFSET

0.00000

5 Press the [Up or Down] key and set

the coefficient for AREA2 (A2).

6 Press the [ENT] key.

The input data is entered and the cursor
returns to CALC.

Note:

When the measurement mode is set to Average Hold, set the averaging
number to 1.