Kistler-Morse KM STXPlus User Manual

Page 38

Advertising
background image

APPENDIX B: KISTLER-MORSE

®

PROTOCOL SERIAL COMMANDS

‘wZ’ Command — Write DSP Filter Qualify
This command is sent by the master to set the DSP Filter qualify of an STXplus. The range of
acceptable values is 2-20. For this example, the master sets the DSP Filter qualify for an STXplus
at address 01 to 15(%). Any of the 5 leading zeros may be omitted.

Request from Master

Response from STXplus

>aawZ00000ddssr

Ar

Format

>01wZ1598r

Ar

Example

‘[R1’ Command — Read 20mA Trim Value
This command is sent by the master to request the STXplus to send the 20mA trim value. The
trim value is used to calibrate the current output engine at the 20mA point. The range of
acceptable values is 0-65535. The default value is 59,674. For this example, the master requests
the 20mA trim value for an STXplus at address 01 which returns 59611 and a checksum of 66.

Request from Master

Response from STXplus

>aa[R1ssr

A00dddddssr

Format

>01[R13Fr

A005961166r

Example

‘[W1’ Command — Write 20mA Trim Value
This command is sent by the master to set the 20mA trim value of an STXplus. The trim value is
used to calibrate the current output engine at the 20mA point. The range of acceptable values is
0-65535. For this example, the master sets the 20mA trim value for an STXplus at address 01 to
55789. Any of the 2 leading zeros may be omitted.

Request from Master

Response from STXplus

>aa[W100000ddssr

Ar

Format

>01[W15578956r

Ar

Example

‘[R2’ Command — Read 4mA Trim Value
This command is sent by the master to request the STXplus to send the 4mA trim value. The trim
value is used to calibrate the current output engine at the 4mA point. The range of acceptable
values is 0-65535. The default value is 11912. For this example, the master requests the 4mA
trim value for an STXplus at address 01 which returns 19611 and a checksum of 62.

Request from Master

Response from STXplus

>aa[R2ssr

A00dddddssr

Format

>01[R240r

A001961162r

Example

‘[W2’ Command — Write 4mA Trim Value
This command is sent by the master to set the 4mA trim value of an STXplus. The trim value is
used to calibrate the current output engine at the 4mA point. The range of acceptable values is
0-65535. For this example, the master sets the 4mA trim value for an STXplus at address 01 to
15789. Any of the 2 leading zeros may be omitted.

B-19