I Troubleshooting

(continued)

Symptom

Possible cause

The mark flashes
when readings are being
taken.

• There was a significant fluctuation in pulse when readings

were being taken. (See page 15.)

The mark flashes
when readings are being
taken.

• Your arm or body moved or the buttons are pressed during

the measurement. (See page 15.)

The mark flashes

even though I did not move

my arm or body.

• In some cases this will be displayed when you have

tensed the muscles in your hand. Relax the muscles in
your hand and take the reading again. (See page 15.)

Appliance is broken.

• Take it to the store.

If the unit still appears to provide unusual or erroneous readings, consult your
physician. If the unit does not appear to be functioning properly, contact Panasonic at
1-800-338-0552.

I Specifications

Power source

DC 6 V (4 /WVLR03 size alkaline batteries)

AC 120 V 60 Hz (when using AC adapter)

Display

Digital LCD

Method of measurement Oscillometric system

Measurement range

Pressure: 0 mmHg to 280 mmHg
Pulse rate: 30 to 160 beats per minute

Accuracy

Pressure: Within ±3 mmHg
Pulse rate: Within ±5 %

Operation temperature/

humidity range

50 °F to 104 °F (10 °C to 40 °C), 15 % to 90 % RH

Storage temperature/

humidity range

-4 °F to 140 °F (-20 °C to 60 °C), 10 % to 95 % RH

Measurable arm

circumference

Approx. 7-3/4" to 15" (20 cm to 38 cm)

Weight

Approx. 7.4 oz. (211 g) (not including batteries)

Dimensions

1 -3/4"

X

6-1/2"

X

2-3/4" (4.5 cm x 16.4 cm x 6.8 cm)

Protection against

electric shock

Internally powered equipment, Type BF applied part [J]

The blood pressure monitor might not perform specifications if stored or used outside the

specified temperature and humidity ranges.