A&D Weighing EJ-1500 User Manual

Page 37

Advertising
background image

35

the unit weight.
Note: Pressing the PRINT key without adding

samples may reduce counting accuracy.

Starting from the 100 samples, 100 - may
be displayed when the sample weight is light.
This is for your notice and press the PRINT

key without adding any samples.

When

aCai 0” (ACAI disabled) or “Umin 2” is

set, this notice is not shown.

CAL errors

Warning to indicate that calibration has been canceled
because the calibration weight is too heavy.

Warning to indicate that calibration has been canceled
because the calibration weight is too light.

Check the weighing pan and the calibration weight.
To return to the weighing mode, press the UNITS key.

Battery error

Warning to show that the batteries are exhausted.
Replace them with new one or use the AC adapter.

Warning to show that the battery voltage is too high.
Check the battery.


AC adapter error

Warning to show that the output voltage of an AC adapter
is too high. Check if the AC adapter is correct.


Warning to show that the output voltage of an AC adapter
is too low. Check if the AC adapter is correct.


Stability error

Warning to indicate that the weight value is not stable
and the balance cannot display it. Prevent vibration and
drafts. Press the UNITS key to return to the weighing
mode.

If you cannot cancel an error or other errors occurred, request service from the
store where you purchased the balance or from your local A&D dealer.










Cal e

-Cal e

lb0

error1

100 -

50 -

Hb1

lb1

Hb0

Advertising
This manual is related to the following products: