7. Deleting the Weighing Data

Note: The AD-1688 performs as follows only when

connected to a weighing instrument.

Two ways of deleting data are available; deleting the most
recent data (or data with the most recent number) or all the
data. Use this to delete the incorrect data.

1. Data number display

Connect the AD-1688 to a weighing
instrument using the RS-232C cable.

When the data number appears,
press and hold the CLEAR key for
2 seconds or more.

2. Selecting a deletion mode.

The “CLR” segment turns on and the most recent data
number flashes.

Under the above condition: Release the CLEAR key to
enter the mode to delete the most recent data.

Keep pressing the CLEAR key to display “ALL”. Release the
CLEAR key here to enter the mode to delete all the data.

If the CLEAR key is kept pressed, the display indicates the
most recent data number and “ALL” alternately.

3. Deleting the data

Press the SET key to delete the designated data. The
display indicates the reduced data number.



If the SET key is not pressed within 5 seconds, the AD-1688
returns to the data number display without deleting the data.

8. Switching the Displays

8.1. Memory Utilization Rate

When the data number is displayed, press the DISPLAY key to
display the memory utilization rate.
Press the DISPLAY key again to display the data number.

8.2. Date/Time Display

When the data number is displayed, press the SET key to
display the year, date and time.

1. Data number display

2. The display indicates the year and date for 2 seconds each,

and then indicates the time.

3. Press any key to return to the data number display.

9. Function Settings

In the function setting mode, settings of the AD-1688 performance
and communication can be changed.
Each function has parameters. By changing the parameter, specify
the performance of the function.
The set parameters are maintained in memory even if the AD-1688
is disconnected from the weighing instrument or personal computer.
For easier setting, the AD-1688 Setting Tools is available for
downloading from our website.

9.1. Setting Procedure

1. Connect the AD-1688 to a weighing instrument using the

RS-232C cable.

2. When the data number appears,

press and hold the SET key for 2
seconds or more.

3. The function display appears. (Refer to “9.2. Function List”.)

Press the DISPLAY key to select a
function, and press the SET key to
display the parameters.

When the CLEAR key is pressed, or
after 10 seconds of non operation,
the AD-1688 returns to the data
number display.

4. The parameter of the selected

function appears. The current setting
has the

mark at the lower right.

The flashing parameter can be
changed.




5. Using the following keys to change the parameter.

DISPLAY (+) key:

Display the parameters one by one.

CLEAR (CANCEL) key: Go to the next function without changes.
SET key:

Confirm the setting and go to the
next function.



After changes, the parameter and the

mark appear for 1 second, and then

the next function appears.

6. Press the CLEAR key to return to the

data number display.

Windows Explorer and Excel are registered trademarks of
Microsoft Corporation.

9.2. Function List

(Note1)

(Note1)

(Note1)

Set these functions according to the settings of the personal computer used.

:

Factory setting

About the number of data on a line and the interval mode, refer to the following sections.

Function

Description

Baud rate

Parameter



600, 1200, 2400*, 4800, 9600, 19200 bps

Data bit, parity

Parameter



7bit-Even*, 7bit-Odd, 8bit-None

Decimal point

Parameter

Dot (.)*, Comma (,)

Number of data on a line

Parameter

1*,2, 3, 4, 5, 6, 7

Buzzer

Parameter

ࠉSounds the buzzer*, Does not sound

Data storage mode

Parameter

Standard mode (All data)*, Interval mode

When the interval mode is selected

Interval time

Parameter

1*,2,5,10,15,20,30,60,120,300,600,900,1200,1800,3600

Date and time data

Parameter Adds the date and time*, Does not add

DATE

SEC

DATE

Date format

Parameter

YMD*, MDY, DMY

DATE

DATE

DATE

SEC

SEC

DATE

SEC

DATE

SEC

9.3. Number of Data on a Line

This function is useful for storing data with different contents or

items at the same time, for example, ID and number of samples
in a counting scale. When the number is set, <CR>s of up to the
setting are converted into commas. Once the number of <CR>s
is reached the setting, a new line starts.

For example, the ID, number of samples and weight are sent

individually. Using this function, these three data items can be
placed on one line as shown in (B).

Example data received to transmit to Excel

ID,000123<CR><LF>

QT,+00000056 PC<CR><LF>

ST,+0003.102 g<CR><LF>

(A) When the number of data on a line is 1, the ID, number of

samples and weight are stored on separate lines.









(B) When the number of data on a line is 3, the ID, number of

samples and weight are stored on one line.

10. Interval Mode

x In the interval mode, the weighing data is stored at a certain interval.
x Available intervals are:

1, 2, 5, 10, 15, 20, 30 seconds / 1, 2, 5, 10, 15, 30 minutes / 1 hour

1. Connect the AD-1688 to a weighing

instrument using the RS-232C cable.

2. Select the interval mode (

int) for the

data storage mode (

reC) of the function

setting and set the interval time (120
seconds in the example).
Press the CLEAR key to display the
data number.

In the interval mode, the

mark turns

on at the lower right.

3. Set the data output mode of the weighing instrument to the

stream mode. For details on the setting procedure, refer to
the instruction manual of the weighing instrument.

4. Press and hold the DISPLAY key for 2

seconds or more. The AD-1688 starts to
store the data at a set interval. The

mark

flashes while the data is being stored.

5. To stop storing the data, press and hold

the DISPLAY key.
The

mark stops flashing and turns on.

6. Remove the AD-1688 from the weighing instrument.

7. Connect the AD-1688 to a personal computer and read the

stored data.

11. Error Display

Error

Description

Remedy



Data full.
No more storage is
available.

Delete the data.

Errors 1 to 3

The voltage that is
supplied from the
weighing instrument
is low.

Use a longer interval
time to store the data.



Error 7

The baud rate and
parity of the received
data is different from
the set values.

Set the same values for
both the AD-1688 and
the weighing instrument.



Error 9

Internal clock error
such as low backup
battery

Ask for repair.



12. Specifications

Devices

Balances and scales with an RS-232C
connector manufactured by A&D
(Visit our website to determine
which devices are applicable.)

Number of data

Approximately 5000 data in the
standard format (15 characters in 1
data set).

Clock error

r1 minute / month

Applicable OS

Windows 2000 / XP / Vista

Dedicated power supply Not required

Power is supplied from a weighing
instrument or a personal computer.

Operating temperature

5°C to 40°C

Operating humidity

85%RH or less, no condensation

Dimensions

55x103x16.5 mm

Mass

Approximately 60 g

No.

CLR

No.

CLR

Mode to delete the

Mode to delete

most recent data

all the data

Release

Press and hold

Release

to be connected

to be stored

Press and hold

No.

No.

Press

Press

From the front

Data number display

Memory utilization rate

Press

No.

%

Press

No.

DATE

DATE

Press and hold

No.

Press

Current setting

No.

ID Number of samples Weight

No.

SEC

No.

No.

Advertising