AD-1683

DC Static Eliminator

Instruction Manual



© 2004 A&D Company Ltd. All rights reserved.

No part of this publication may be reproduced,
transmitted, transcribed, or translated into any language
in any form by any means without the written permission
of A&D Company Ltd.

The contents of this manual and the specifications of the
instrument covered by this manual are subject to change
for improvement without notice.




To Use the AD-1683 Safely

Thank you for your purchasing the AD-1683 DC static eliminator.

Please read this manual completely before using the AD-1683.

Warning Definition

The warning described in this manual has the following meaning:

Caution

Important information to alert you to a situation

that may cause injury and/or damage to your

property if instructions are not followed.

When operating this instrument, be sure to observe the

following safety precautions

Caution

Do not open the case to repair the AD-1683. Only qualified

personnel can do that. Attempting repairs yourself may cause

damage to the AD-1683. Damage caused by attempting to do

the repair yourself will void the warranty.

Do not use the AD-1683 in a closed environment. Ozone

that is harmful to the human body may stray into the air.

Ventilate well.

The AD-1683 is not explosion-protected. Do not use the

instrument where flammable solvents or dust exits.

Do not touch the discharge electrode pins. It may cause an

electrical shock.

Use the specified AC adapter only. Please confirm that the

AC adapter type is correct for your local voltage and

receptacle type. If an AC adapter other than specified is

used, static electricity may not be eliminated or it may

cause an electrical shock. (*1)

Be sure to insert the AC adapter into a grounding-type

electrical outlet and earth ground the grounding cable.

Otherwise, static electricity may not be eliminated or it may

cause an electrical shock.

*1: Use the AC adapter identification label to identify the AC adapter.

Caution

Do not keep the AC adapter connected to the electrical

outlet for a long period of time. Dust that has collected on

the outlet may cause a fire.

When relocating the AD-1683, remove the AC to avoid a

possible electrical shock.

Precautions on Usage

Caution

If the AD-1683 malfunctions, stop using it immediately. Place

an “Out of Order” label on it or move the instrument to a

place where it can not be used by mistake. It is very

dangerous to continue using the instrument under such

conditions. Ask the nearest dealer for services.

Do not face the front of the AD-1683 directly toward an

electronic balance. Ionized air from the AD-1683 may affect

the measurement and cause a measurement error.

When used with a balance of other manufacturers, static

elimination can not be guaranteed.

Do not place the sample closer to the AD-1683 than specified.

The sample may become charged.

Do not place any obstacle between the AD-1683 and the

sample.

The stand can be attached to the workbench. The stand has

three holes with a diameter of 6 mm in 42.5 mm interval for

that purpose.

(To secure the stand: Bend the side plates outward to remove

the AD-1683, using much care not to separate the upper and

lower cases. Attach the stand to the workbench using the

screws. Replace the AD-1683 on the stand.)


1. Introduction

This manual describes how the AD-1683 works and how to get the

most out of it in terms of performance. Please read this manual

completely before using the AD-1683.

2. Features

The AD-1683 generates bipolar ions continuously by DC corona

discharge separately from positive and negative discharge

electrodes and directs the ionized air onto the charged body to

eliminate static electricity. The generated ions are well-balanced

in polarity and can eliminate static electricity regardless of the

polarity of the charged body. (See Fig.3.)

Static electricity

Generally, insulators such as plastics are prone to be charged

with static electricity when the ambient relative humidity is

below 45%RH. This may affect weighing and cause a

measurement error of several milligrams.

The AD-1683 can eliminate static electricity very effectively.

3. Part Names
















Note: Please confirm that the AC adapter type is correct for your

local voltage and receptacle type.

To prevent electrical shock, insert the POWER cable into a

grounding-type outlet and earth ground the grounding cable.

4. How to Use the AD-1683

4-1. Installing the AD-1683

Choose the installation site so

that the AD-1683 can be placed

beside the electronic balance

and a space wide enough to

place a sample (10 cm to 30 cm

from the front of the AD-1683)

can be secured. Place the

sample in front of the AD-1683

inside the specified area and

perform static elimination. Then,

weigh the sample using the

electronic balance.


























4-2. Turning the power on

Insert the AC adapter plug to the AC adapter jack located on the

rear of the AD-1683. Insert the AD adapter into an electrical

outlet. The red LED at the center of the AD-1683 illuminates to

indicate that static elimination is available.

Note: Please confirm that the AC adapter type is correct for

your local voltage and receptacle type..

4-3. Finishing the use

Remove the AD adapter from the electrical outlet.

5. Maintenance

5-1.Cleaning

Caution

Remove the AC adapter from the AD-1683 before cleaning.

When used for a long period of

time, dust may collect around the

discharge electrode pins and

static elimination performance will

be lowered. To maintain the

elimination performance, clean the

AD-1683 periodically (once a

week). Use a cotton swab to clean

the electrode pins.







5-2. Replacing the discharge electrode pin

When the elimination performance

remains low even after cleaning

the discharge electrode pins,

replace the discharge electrode

pins with new ones.

Remove the AC adapter from the

receptacle and the AD-1683.

Using a pair of pliers, pull the

discharge electrode pin out.

Insert a new pin, using much care

not to bend the pin tip. Be sure to

insert the pin till stopped.


Replacement discharge electrode pins (Sold separately):
AX-054016580-S

6. Specifications

Static elimination method

DC corona discharge (Bipolar)

Elimination range

(See Fig. 2)

Distance: Approx. 10 cm to 30 cm from the electrode

Area : 30x30 cm from the electrode

Elimination performance

When charged 5 kV

Distance 10cm

20cm 30cm

50cm

Elimination time

1 sec

4 sec 15 sec 70 sec

Ambient temperature and humidity 0°C to 40°C, 80%RH or less
Ozone concentration

Positive discharge pin: 0.07PPM

Negative discharge pin: 0.25PPM

(Measured at a distance of 1 cm from each discharge pin.)

Power supply

AC adapter. Please confirm that the AC adapter type
is correct for your local voltage and receptacle type.

Power consumption

Approx. 1.5 VA

Discharge electrode pin

Tungsten (

φ

0.9x17 mm), Life Approx. 10000 hours

Mass

Approx. 300 g (including the stand)









External Dimensions

















WM+PD4000962