Ventalair, Important safeguards, Instruction for use – Allersearch Ventalair Max User Manual

Page 2

Advertising
background image

Ventalair

®

MAX

Product Code No. AE 650-044

Congratulations on purchasing your Allersearch nebuliser system.

The Ventalair

®

MAX is the latest model in the Allersearch range of nebuliser therapy systems and is suitable

for use by more than one person in the family. The Ventalair

®

Max comes complete with mask set suitable

for child or adult.

IMPORTANT SAFEGUARDS

The Ventalair

®

MAX is a medical device for home use and is only to be used with medication

prescribed or recommended by a doctor and in accordance with the following operating instructions.
It is important that the patient reads and understands the information provided for use and care of the
unit. Please contact your pharmacy or Allersearch if further information is required.
The Ventalair

®

MAX is manufactured to the highest quality and to comply with safety standards. However

as with any electrical product, the following basic precautions should be observed when using the
Ventalair

®

in order to avoid risk of personal injury or damage to the product. Before using the Ventalair

®

MAX for the first time and periodically there after check the power cable for possible damage. Any sign
of damage or wear and tear immediately disconnect from electric socket and return the unit at were
initially purchased.
• Children and the disabled should use the unit under the strict supervision of a person who has read

this manual.

• The unit must not be used in the presence of inflammable gases such as oxygen nitrous oxide or

anaesthetic substances.

• Never use adaptors where power supply is different to the one stated on the identification plate. Keep

the cable away from heated surfaces.

Do not handle the electrical plug with wet hands. Never use the unit while having bath or shower.
Do not submerge the unit in water, should this accidentally occur immediately switch off the power

and unplug cable from electric socket. DO NOT REACH FOR THE SUBMERGED UNIT until the electric
cable is unplugged from the electric socket. Do not use unit until it has been serviced by an Allersearch
authorised service agent.

• The Ventalair

®

MAX case is not protected against the ingress of liquid.

• Unplug the unit after each use.
• The Ventalair

®

MAX is fitted with a non-replaceable thermal fuse to protect the user if motor over

heating should occur.

• Keep a spare jet nebuliser bowl handy in case the other one is lost or damaged.
• Do not block air vents or place on soft surfaces such as beds or carpet as the air vents may become

blocked and reduce the cooling effect on the motor.

INSTRUCTION FOR USE

After each therapy session, the jet nebuliser bowl and accessories should be cleaned according to the
instructions in the “CLEANING and DISINFECTION” section.
1. Connect the power cord (6) into the socket (5) and plug the power cord to the electrical socket.
2. Wash hands before preparing the medication.
3. Insert the elastic into both sides of the face mask.
4. Unscrew jet nebuliser bowl (8) and pour the prescribed medication into the bottom section (8a)
5. Reassemble the top section (8c) to the bottom section (8a), connect air supply tube (7) to the bottom

of jet nebuliser bowl (8a) and the air outlet port (4).

6. Fit the appropriate size mask to the jet nebuliser bowl (8) and place on face. If mouthpiece (11) is to

be used fit this in place of mask.

7. Sit in comfortable position, press switch (2) and breathe normally through the mouth. To obtain maximum

medication, occasionally tap the jet nebuliser bowl.

8. When mist stops, switch off unit and unplug from electrical socket. Allow to cool down before storing

away.

CLEANING and DISINFECTION: Wash hands before cleaning or disinfecting the accessories. All components
should be rinsed under tap water after each use. Clean the centre jet (8b) after detaching it from the
top section (8c) of the jet nebuliser bowl and disassembling as shown in fig. A. All components, except
air tube and mask, can be disinfected by boiling for 5-10 minutes. Alternatively all components, except
the air tube, can be chemically disinfected by using a mild cleansing solution such as Allersearch
Nebuclean available from pharmacy – follow the disinfecting instruction and then rinse components
under tap water. Place components on paper towel and let air dry. Do not wipe components with towel
or tissue as their particles may block the jet nebuliser bowl causing poor misting. Reassemble the centre
jet (8b) and fit it back in place on the jet nebuliser bowl as shown in fig. A.

Do not use abrasive cleaners, immerse or run water over the Ventalair

®

MAX case. Regularly wipe case

with damp cloth and mild detergent.

NOTE: Replace your jet nebuliser bowl every 6 months if it is used regularly. This will ensure that your

medication is always delivered at the correct particle size.