ALTANA BYK-mac Short Instructions User Manual

English, Operation, How to calibrate

Advertising
background image

269 020 584 E 0806

Additives & Instruments

A member of

Measure what you see.

BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments

Please note detailed operating instructions in English on CD-ROM.

ENGLISH

Operation

mode wheel:

s

Switch

on

press

s

Show

menu

press

s

Select

function scroll

s

Activate

function press

esc button:

s

Change between menu

and measurement mode

std button:

s

Short-cut to measure standard

Operate button:

s

Switch on, select and measure

How to Calibrate?

Black/white calibration

s

Select Setup – Calibrate – Black/White Calibration

s

Place unit on black trap and confirm OK

s

Place unit on white standard and confirm OK

Black/white calibration is requested every 3 months.

Cyan check
After a black and white calibration the instrument must be checked on the
Cyan standard.

s

Place unit on Cyan standard and confirm OK

The Cyan check is requested every 7 days.

operate

How to measure Differences?

s

Select a stored standard or measure a std.

s

Input standard name:

Change character ==> turn scroll wheel

Move to next digit ==> press scroll wheel

Move one digit back ==> press esc

s

Input number of readings

s

Place instrument on standard

s

Press operate to take a reading

s

Copy predefined settings for data display

s

Continue with sample readings

Note: All color data inclusive spectral data are
saved and transferred to software.

s

First page displays color

differences

s

Scroll to second page for sparkle

and graininess results