13

Symbols used in the Display

∞ Fe

measurement on steel or iron substrate

∞ NFe measurement on non-magnetic substrate
∞ Error handling

error

∞ Infi

measurement beyond range, wrong substrate

∞ !

upper limit exceeded

∞ "

lower limit exceeded

∞ x average
∞ # standard

deviation

∞ $

wireless interface activated

continuous measuring mode

Buzzer switched off

batteries week, measurements can still be taken

batteries

empty

14

Scope of Supply

∞ Gauge with optional probe

∞ Adapter cable for probe

∞ 2 batteries AA 1,5V (alkaline)

∞ Instruction

manual

∞ Test certificate for optional probe

∞ Soft pouch with clip

∞ Case for transport and storage
Optionally available
∞ Fe-Probe 2 mm (80 mil)

∞ Fe-Probe 5 mm (200 mil)

∞ NFe-Probe 2 mm (80 mil)

∞ Dual-Probe Fe/NFe

2 / 2 mm (80 / 80 mil)

∞ Dual-Probe Fe/NFe

5 / 2 mm (200 / 80 mil)

∞ Calibration shims, Zero reference plates, also available as set

QuaNix

®

-Software for transfer to and evaluation of the data in EXCEL and gauge

configuration (*)

∞ Wireless interface for data transfer between gauge and PC (*)

(*) Already included in the scope of supply of the byko-test 8500 Premium version



7.3.4.3

Information About a Batch

The item "INFO" returns information about date and time when the batch was created, and
how many readings it contains.

7.3.4.4 Limits

Upon starting this function the gauge displays the limits stored with the batch. If no limits are
set in the batch, the gauge displays the last reading.
Upper and lower limits can be set for each batch individually using the same procedure as
described in §§ 7.2.3 and 7.2.4.
The modified limits are used in the active batch instead of the global limits.

7.3.4.5

Rename a Batch

Renaming a batch is done in the same way as described in § 7.3.3.

7.3.4.5

Delete a Batch

The entire batch can be deleted when calling the delete function and finally confirming the
procedure by pressing the key

%.

7.4 Statistics

The gauge calculates the following parameters:

- Average
- Standard

deviation

- Maximum
- Minimum

Within the menu for the statistics two display modes are used:

a)

In the memory mode two statistical parameters of the active batch can be
displayed. These parameters are updated with each new reading added to the
batch.

b) When

calling the statistics of a non-active batch all parameters are displayed.

When entering this menu, the list of all batches is displayed.

7.4.1

Selection of Statistical Parameters of the Active Batch to be

Displayed

After selecting the active batch and pressing key

% the list of statistical parameters is

displayed. With the keys

! and " the individual parameters can be selected and be

activated by pressing key

%.

NOTE: As a maximum 2 parameters can be activated to be displayed !
To leave this menu item press the (backspace-) key

&. For an immediate return to the

normal display mode just press the probe tip shortly.

7.4.2

Display of Statistics of not activated Batches

When selecting a non-active batch using keys

!and " and confirming the selection with

key

% all statistical parameters of this batch are displayed in a popup-window.

7.5 Gauge-Setup

7.5.1 LC-Display

7.5.1.1 Unit

of

Readings

The readings can be displayed either in

'm or in mil (standard setting: 'm).