American Diagnostic Corporation (ADC) ADC Proper Maintenance User Manual

Just in case, Adc guidelines

Advertising
background image

JUST IN CASE . . .

ADC GUIDELINES

for the Proper

Maintenance,

Handling, Cleanup,

and

Transportation

of

Mercurial Blood

Pressure Devices

PROBLEM

LIKELY CAUSE

RESOLUTION

Air Bubble in

• Insufficient amount

• Add mercury by following the instructions

cartridge tube

of mercury

in this manual

during inflation

• Air trapped in

• To remove trapped air, tap reservoir while

reservoir

system is inflated

Blood Pressure

• Excessive deflation

• Be sure to deflate the inflation system at

Reading seems High

rate

the recommended rate of 2-3mmHg/sec

• Air flow blockage

• Change top cap filter

• Clean or replace mercury

Blood Pressure

• Mercury level low

• Add mercury

Reading seems Low

Dirty cartridge tube

• Mercury has oxidized

• Clean tube in accordance with enclosed

leaving a silver gray

instructions

residue

Mercury bouncing in

• Filters need

• Change top cap filter

tube during inflation

replacements

or deflation

Mercury drops

• Leak at seals

• Make sure cartridge tube is securely seated

appear outside the

• Replace silicone gaskets at top and bottom

top or bottom of the

of tube

cartridge tube

Mercury rises or

• Air flow blockage

• Change top cap filter

falls slowly

• Clean tube in accordance with enclosed

instructions

Mercury separates in

• Excessive deflation

• Be sure to deflate the inflation system at

tube during deflation

rate

the recommended rate of 2-3mmHg/sec

• Trapped air

• To remove trapped air, tap reservoir while

system is inflated

IB p/n: 9353B-00 - rev 4

To Register Your Product, visit us

at www.adctoday.com

and follow the links

Consult this manual for basic service

and maintenance procedures mentioned

in this troubleshooting guide.

If any of the problems persist, contact

ADC’s Customer Service Department

Toll-Free at 1-800-ADC-2670.

Advertising