5. With your index finger on the rubber focusing ring, adjust for optimal focus while

your eye is resting on the rubber eye guard.

6. Observe the illuminated and magnified lesions, making notes for treatment.

7. To measure pigmented skin lesions, the graduated contact plate must be used.

8. Switch off the unit after use to preserve batteries. Twist the black knob counter-clockwise

until an audible click is heard and the button comes to its off position.

Care and Maintenance:

NOTE: Periodically check the battery conditions, making sure that no sign of corrosion

or oxidation is present.

Cleaning and Sterilization:

1. Contact Plate (Fig. 2).

The contact plate must be taken off after each examina-

tion.

The contact plate can withstand cleaning or disinfection

by soaking in alcohol solution, cidex, etc.

The contact plate can also be autoclaved up to 135°C and

pressure of 28 p.s.i.

NOTE: Clean the contact plate with a cleaning solution prior to autoclaving.

2. Lens

The lens can be cleaned by using a moistened cotton-tipped applicator with alcohol or

a solution of water with mild detergent.

Do not use abrasive materials to clean the lenses as they will scratch the coated sur-

face of the lens.

3. Handle

The handle can be cleaned with a moist cloth.

Battery handle shell should not be emersed in water. Remove battery components

while cleaning.

Lamp Replacement:

NOTE: (Fig. 2).Before removing the lamp, make sure that the device

has been turned off for some minutes to avoid the risk of being burnt.

Lamps may be hot, so it’s necessary to let it cool.

Remove Dermascope head from the power source as advised in

operating instructions (Fig. 1).

Pull out lamp with nail file or similar tool (if necessary) by getting

it under the base of the lamp (Fig 3),

Use an ADC recommended lamp for replacement by carefully

aligning pin on lamp shaft with slot in Dermascope head. Push

the lamp until sealed well.

Battery Replacement:

NOTE: Handle batteries carefully when removing, as the liquid in them can irritate the skin

and eyes. Dispose of the used batteries properly. Replace with alkaline type “C” batteries.

(Fig. 2)