American Diagnostic Corporation (ADC) Satin Standard User Manual

Page 2

Advertising
background image

Thank you for choosing an ADC® Satin™ Laryngoscope.

Device Description and Intended Use

A Laryngoscope is a two-part, hand held device consisting of a handle that contains batteries and a de-

tachable blade. Laryngoscopes are designed to provide illumination within the larynx during the process

of performing intubations. This instruction book covers both standard, fiberoptic type and disposable

laryngoscopes. Fiberoptic Laryngoscopes are indicated by an F following the item number / Disposables

are indicated by a D following the item number.

Contraindications

The following are only relative contraindication to tracheal intubation:

1. Severe airway trauma or obstruction that does not permit safe passage of an endotracheal tube.

Emergency cricothyrotomy is indicated in such cases.

2. Cervical spine injury, in which the need for complete immobilization of the cervical spine makes en-

dotracheal intubation difficult.

General Warnings

A warning statement in this manual identifies a condition or practice which, if not cor-

rected or discontinued immediately could lead to patient injury, illness, or death.

WARNING: Federal law restricts this device to sale by or on the order of a physician or licensed healthcare

practitioner.

WARNING: ONLY TRAINED PERSONNEL SHOULD USE A LARYNGOSCOPE FOR INTUBATION

WARNING: ADC

®

recommends that spare lamps, batteries, and replacement parts always be available for emer-

gency use.

WARNING: The sterilization guidelines included in this instruction booklet, provided by American Di-

agnostic Corp. (ADC

®

), are intended as procedures compatible with specific materials. Sterilization

must be performed to approved hospital protocol. ADC

®

cannot guarantee that any of the recommended

methods will guarantee sterility. This must be validated by the hospital and/or sterilization equipment

manufacturer.

WARNING: Fiberoptic handles may be used with ALL fiberoptic blades (reusable or disposable) that

conform to the ISO 7376-1 and ISO 7376-3 standards. Standard handles may be used with any standard

blades that conform to the ISO 7376-1 and ISO 7376-2 standards. Standard and fiberoptic blades and handles

are NOT interchangeable.

Operation of all Laryngoscope Models

To Activate: Attach selected blade onto handle by hooking blade onto

handle rod as shown in Figure 1. Pull blade upward to open position to

automatically activate light. To turn off, push blade downward into off

position and disconnect from handle. (See Figure 1).

Compatibility/Standards

Standard ADC

®

Satin™ Laryngoscopes conform to current ASTM F965, ISO 7376-1, and ISO 7376-

Figure 1

Advertising