proper placement of the probe tip could result in a longer time for a reading.

Rectal use:

1. Commonly used for babies, young children or when it is difficult to take an oral or underarm tem-

perature.

2. After placing the probe cover on the thermometer, lubricate with a water-soluble jelly or petroleum jelly .

3. Gently insert the probe tip no more than 3/4 inch or 2-3 cm into the rectum. Do not force the

tip into the rectum if resistance is encountered.

4. Dispose of the used probe cover and wash the thermometer as recommended.

5. The peak temperature should be reached in approximately 10 seconds.

Underarm (Axillary) use:

1. Wipe underarm with a dry towel.

2. Place probe tip in armpit and bring arm down against the body.

3. The probe should be pointing upward in the same direction as your body if possible.

4. The time required for an accurate underarm temperature may be 20 seconds or longer depending

on placement of the probe and closure of the arm against the body.

Changing Measurement Scale: From the OFF position, press the Multi-Function button to acti-

vate the unit. Immediately after releasing the Multi-Function button, press the Multi-Function button

again and hold for 2 seconds. Press the Multi-Funtion button again to change from °F to °C or °C to

°F, and toggle back and forth. Release the Multi-Function button to set.

Memory Recall:

Press the Multi-Function button for approximately 5 seconds. The last temperature will be recalled

and displayed as long as the button is held down. When the button is released, the thermometer will

be in “ready-to-measure” mode and the recalled temperature will be erased.

Changing the battery:

When the “

” appears in the lower right of the display, the battery is exhausted and needs replacing.

Make sure a battery of the same type is on hand and proceed as follows: Remove battery compartment lid.

Use a small pointed object, but not sharp metal. Withdraw the battery. Replace with a new button battery

(1.55v, LR41 or SR41) with the Cathode side (-) upward. Carefully replace the battery compartment lid se-

curely. Properly dispose of the batteries, keeping them from small children and heat.

Standards:

Complies with EN 12470-3:2000 and ASTM E 1112-00

How to care for this product:

• Use of probe covers is recommended to help avoid spreading germs, and to keep the thermometer

clean. The thermometer should be cleaned after use even when probe covers are used.

• Clean the thermometer tip and probe only using isopropyl alcohol or soap and water.

• Store the unit in the protective case while not in use.

• Do not store or keep in direct sunlight, or in high temperatures.

• Keep battery away from children.

• Do not open the thermometer except to replace the battery.

• Do not drop the thermometer, as this may ruin the unit.

• The manufacturer recommends verifying the accuracy by an authorized laboratory every 2 years.