American Diagnostic Corporation (ADC) Adsafe CPR 4053 User Manual

Page 4

Advertising
background image

Cleaning:

Mask is reusable and easy to clean. One-way valve can be cleaned and reused ONLY for re-

peated mannequin training. Use provided alcohol wipe to clean mask and valve after use

Discard filter after each use.

Wash and scrub in warm soapy water. Rinse in clean water.Submerge for 10 minutes in a

1:64 household bleach:water solution. Rinse again and allow to dry.

Do not pasteurize, boil, gas, or steam autoclave mask.

Note:

If mask becomes sticky, clean as recommended above.

Use provided alcohol wipe to clean mask and valve after use (with included models).

Storage:

Always keep mask in storage case when not in use.

SPECIFICATIONS

Dimensions: 5.1” x 4.2” x 1.9”

(130mm x 106mm x 48mm)

Operation Temperature: -0.4°F ~ 104°F (-18°C ~ 40°C)

Storage Temperature: -22°F ~ 122°F (-30°C ~ 50°C)

Storage relative humidity: <85%

Operation Function: Inspiration/Expiration resistance

less than 1.8” H2O at 50 LPM

Regulatory Standards:

ISO 10651-4: Lung Ventilator

Part 4: Particular requirement for operator powered resuscitators

Latex-free

Valve and Gloves are single use only.

ADC
55 Commerce Drive

Hauppauge, NY 11788

U.S.A.

ADC (UK) Ltd.
Unit 6, PO14 1TH
United Kingdom

Inspected in the U.S.A.

Made in China

tel: 631-273-9600, 1-800-232-2670

fax: 631-273-9659

www.adctoday.com

email: [email protected]

IB p/n 93-4053-00 rev 2 english