Testing – AXEON Fluid-O-Tech Repair Rebuild User Manual

Page 14

Advertising
background image

08/09 Ed., Rev. B

Page 14of 16

Testing

It is recommended that the pump be tested before being put into service, to ensure that the

pump was assembled properly and is running satisfactorily. The pump can be tested as

follows:

1) Connect the pump to a motor using the proper coupling and adapter if necessary (contact

Fluid-o-Tech for details, or visit

www.fluidotech.com

). Plumb the inlet of the pump directly to a

source of clean water, using tubing with an ID sufficiently sized for the flowrate of the pump.

Plumb the outlet of the pump to an inline pressure gauge, then to a ball valve, then to a flow

meter if one is available, then back to the reservoir of water (or drain if you are using a city

line). Be sure that no teflon tape or pipe dope enters the lines, as this can cause significant

damage to the pumps.

2) Turn the unit on and allow the pump to prime and rid the circuit of air. This shouldn’t take

too long, but if after a long period of time air still remains in the lines, check the following: the

fittings are sealed in the ports properly, the bypass is screwed on tightly and has a nylon

washer, the hoses in the tank are well below the surface of the water and not too close to

each other, and that there are no other leaks in the circuit. If the pump at any point begins to

make a buzzing noise, this is usually a sign of cavitation. To remedy, increase the inside

diameter of the inlet hose and/or shorten it’s length.

3) Once the pump is up and running, check for any obvious leaks. The important points to

examine are the cap, weep holes, bypass, and inlet and outlet. It is normal, on occasion, for a

few drops of liquid to come out of the weep holes on initial startup, especially if there is

residual ethyl alcohol in the cavity. This isn’t a problem unless it is a large amount or a

continuous drip. Next, take a reading on the flow meter. It should be a value close to the

flowrate listed on the corresponding data sheet for the particular pump and speed you are

using. Now close of the valve in the line to raise the pressure to about 100 psi. Re-check for

leaks and take another flow reading. Allow the pump to run for five or ten minutes. If after

Advertising