0 maintenance and service, 1 cleaning, Figure 2 – Boekel Scientific Analog Incubators User Manual

Page 10: Fuse drawer

Advertising
background image

Page 10 of 12

8.0 Maintenance and Service


All Boekel laboratory products are designed to comply with IEC1010-1.
No routine maintenance is required.
8.1

Cleaning

Disengage power cord prior to cleaning. If a spill occurs, use appropriate
clean-up procedures as required for radiation or biohazard control. The
outer casing can be cleaned with a cloth dampened with water. Do not
immerse the incubator in water.
8.2

Replacement of Fuses (all models except 115 V 5.0 Cu. Ft.)

Caution – Double pole / Neutral Fusing

There are two supply fuses located in the fuse drawer. To change the
fuses:
• Turn power switch to the off (O) position

• Disconnect the unit from the power supply

• Remove the line cord from the power entry module on the back of the

unit

• Pull back on the fuse drawer catch

• Pull out the fuse drawer (See Figure 2)

• Check and replace with the correct fuses if necessary. The fuses

must be 5mm x 20mm quick acting, rated at 250V.

0.5 Cu. Ft. (115VAC)

2.5AF

0.5 Cu. Ft. (230VAC)

1.25AF

1.4 Cu. Ft. (115VAC)

2.5AF

1.4 Cu. Ft. (230VAC)

1.25AF

2.5 Cu. Ft. (115VAC)

2.5AF

2.5 Cu. Ft. (230VAC)

1.25

5.0 Cu. Ft. (230VAC)

3.15AF

• Push the fuse drawer back in. Reconnect unit to power supply.

Fuse
Drawer

Figure 2