Buffalo Filter PlumePen Elite User Manual

Plumepen, Elite, Surgical plume evacuation pencil

Advertising
background image

If you have any questions or need any assistance with the set-up

or use of the PlumePen

®

Elite

Surgical Plume Evacuation Pencil

please contact Buffalo Filter

®

at:

U.S. toll-free: 800.343.2324

International: 1.716.835.7000

Email: [email protected]

5900 Genesee Street, Lancaster, NY 14086

U.S. toll-free 800.343.2324 . International: 1.716.835.7000

Fax: 716.835.3414 . Email: [email protected]

www.buffalofilter.com

PlumePen

®

Elite

Surgical Plume Evacuation Pencil

CAUTION:

DO NOT USE in patients who have electronic implants such as cardiac

pacemakers without first consulting a qualified professional (e.g.,

cardiologist). A possible hazard exists because interference with the action

of the electronic implant may occur, or the implant may be damaged.
INSPECT instruments and cables for damage prior to each use, especially

the insulation of laparoscopic/endoscopic instruments. This may be done

visually under magnification or with a high voltage insulation testing

device. Insulation failures may result in burns or other injuries to the patient or operator.
Sterility unless packaging is damaged or any seal is broken. Do not use electrosurgery

in the presence of flammable anesthetics or other flammable gases, fluids, or objects,

or in the presence of oxidizing agents, as fire may result. Place active accessories in a

holster or in a clean, dry, non-conductive and highly visible area away from the patient

when not in use. Inadvertent contact with the patient may result in burns. Contact with

drapes or linens may cause a fire.
DO NOT activate the instrument when not in contact with target tissue, as this may cause

injuries due to capacitive coupling.
PlumePen

®

Elite is designed and intended only to be used with an electrosurgical

generator that has been tested to the IEC 60601 standard. Please refer to generator to

ensure compatibility. PlumePen

®

Elite is not a fluid removal device, therefore, it should

not be used for such application.
ASPIRATE fluid from the area before activating the instrument. Conductive fluids (e.g.,

blood or saline) in direct contact with or in close proximity to an active electrode may

carry electrical current or heat away from target tissues, which may cause unintended

burns to the patient.
Buffalo Filter recommends that the original blade be utilized with the PlumePen Elite.
If original blade is removed, visually confirm new blade is completely inserted and

secured before activating the pencil. Never force the blade into the pencil. Rotate the

pencil if the smoke tube is obstructing view of the insertion point.
For procedures where visualization may be impaired, be alert of these potential hazards:
• The electrode tip may remain hot enough to cause burns after the current has been

de-activated.

• Inadvertent activation or movement of the activated electrode outside of the field of

vision may result in injury to the patient.

• Localized burns to the patient or physician may result from electrical currents carried

through conductive objects. Current may be generated in conductive objects by direct

contact with the active electrode, or by the active accessory being in close proximity

to the conductive object.

For single use only. Do not reuse, reprocess, or re-sterilize. Reuse, reprocessing, or

re-sterilization may compromise the structural integrity of the device and/or lead to device

failure which in turn may result in patient injury, illness, or death.
This single-use product is not designed or validated to be reused. Reuse may cause a risk

of cross-contamination or cause a malfunction as a result of the product being physically

damaged due to cleaning, disinfection, re-sterilization, or reuse.

Indications For Use: The PlumePen

®

Elite is designed for general electrosurgical

applications, including cutting and coagulation, and for removing smoke generated

by electrosurgery when used in conjunction with an effective smoke evacuation

system. The pencil enables the operator to remotely conduct an electrosurgical

current from the output connector of an electrosurgical unit to the operative site for

the desired surgical effect.
Indications for Use for the PlumePen

®

Elite Integrated Smoke Evacuation Pencil

include:
a. To remove smoke plume from the surgical site.
b. To remotely conduct an electrosurgical current from the output connector of an

electrosurgical unit to the target tissue for the desired surgical effect.

Figure 4.

The PlumePen

®

Elite is equipped with several key features for the

surgeon’s convenience. Above the electrode blade is a translucent

plastic tube that can be positioned at varying lengths, (Figure 4) to

most effectively capture surgical smoke plume as it is created.
With the electrosurgical generator off and the electrode blade

cool, position this tube as near the point of the tissue interaction

as possible without obstructing the view of the target tissue.
Keep the active electrodes clean. Build-up of eschar may reduce

the instrument’s effectiveness. Do not activate the instrument

while cleaning. Injury to operating room personnel may result.

Figure 5.

The PlumePen

®

Elite has CUT and

COAG buttons located on the top of the

electrosurgical pencil body. The YELLOW

button is for cutting and the BLUE button is

for coagulation, (Figure 5).
After use, turn off smoke evacuation unit and

electrosurgical generator. Discard the entire

PlumePen

®

Elite assembly (Pencil, Blade,

Tubing, and Power Cord) per facility protocol

for contaminated items.

Figure 6.

To optimize the life of your smoke evacuation filter,

it is recommended that a Buffalo Filter Surgical

Smoke Plume Evacuator be used in conjunction with

an EZLink™ Automatic Activation Device (Model

EZLink01), (Figure 6).
The EZLink™ will automatically turn the Buffalo Filter

Surgical Smoke Plume Evacuator on or turn it off

when either the CUT or COAG function is activated/

de-activated on the PlumePen

®

Elite . Contact Buffalo

Filter

®

for more details on the EZLink™

.

904162 RevB.pdf

Figure 1.

Open product packaging and remove the

PlumePen

®

Elite Surgical Plume Evacuation

Pencil assembly. Connect the tubing provided

with the PlumePen

®

Elite to a smoke evacuation

system or other appropriate vacuum source;

example shown, (Figure 1). For connection and

operation to a specific smoke evacuation system,

please refer to the specific instructions for use for

appropriate setup and use.

Figure 2.

Prior to connecting the PlumePen

®

Elite 3-prong

power cord to the electro-surgical generator,

ensure that the electrode blade (provided) is fully

inserted into the pencil by gently pressing it in

front of the PlumePen

®

Elite . Remove the plastic

shield from the blade and discard.

Figure 3.

Plug the 3-prong power cord into the electrosurgical generator

of your choice, (Figure 3). Confirm that all power settings on the

generator are appropriate for the procedure being performed.

The electrosurgical generator’s intensity should be set as low

as isnecessary to achieve the desired effect. The PlumePen

®

Elite

is a monopolar electrode, the use of a dispersive electrode

is required to prevent burns/injury to patient. Please refer to

electrosurgical generator user manual and dispersive electrode

instructions for use for additional instructions.

ONLY

CAUTION:

Intended for use with a

maximum voltage of 10kV P-P.

Do Not Use

Damaged

Single Use Only

Do Not Reuse

PLP1001

PLP2001

PLP1501

PLP2501

Do Not

Resterilize

STERILE R

Does Not

Contain Natural

Rubber Latex