Flintec LCT-11 User Manual

Lct-11 quick user guide

Advertising
background image

LCT-11 Quick User Guide V.1.00, September 2012

Page 1 of 3

LCT-11 Quick User Guide


1. Before the instrument can be used for the first time you have to insert 4 batteries of type AA. The battery

box can be accessed from the rear side of the instrument after the protective cover has been removed.
After the batteries have been inserted you can refit the protective cover.

2. The Instrument is off. If this is not the case please turn off the instrument by pressing the <On/Off> key for

approx. 5 seconds.

3. Connect the wires of the oad cell cable to the correct pins of the input connector, but make sure that the

SCREEN and the BODY are NOT connected.

4. Power on the instrument by pressing the <On/Off> key for approx. 2 seconds.
5. The display will show the “Welcome” screen for a few seconds. Then it will ask for the load cell connection

type: 4-wire OR 6-wire.

6. For the best accuracy of the high resistance measurements (leakage) it’s now a good opportunity to run a

calibration. Press the <

> and the <On/Off> key simultaneously. The display will show “Calibrating”...and

after 2 seconds the instrument is calibrated.

7. Now connect the SCREEN and the BODY (of the load cell) to the input connector.
8. Select your load cell connection type (4-wire or 6-wire) and press <Enter>.
9. Select the rated output of the load cell with the <

> and <

> keys and press <Enter>.

10. Press the <Enter> button for a full load cell test cycle OR press the <

> key for the continuous display

mode (GAIN).


If the Full load cell test cycle was selected:

The LCD will display “Test in progress” and the LED will flash for a few seconds. Once the test cycle is
finished you may scroll the results by pressing the <

> key.

The results will be displayed in the following sequence:

Input

Resistance between Ex+ and Ex- (Max. 5000 ohm)

Output

Resistance between Signal+ and Signal- (Max. 5000 ohm)

Sense+

Resistance between Ex+ and Sense+ (Max. 500 ohm; valid for 6-wire cables only)

Sense-

Resistance between Ex- and Sense- (Max. 500 ohm; valid for 6-wire cables only)

Signal output

Load cell signal in percentage of full load cell capacity

Note: In case of a broken cable or a very high input/output resistance the display
will show “Can not measure”

Shield to bridge

Insulation resistance between the cable shield (screen) and the bridge

Body to bridge

Insulation resistance between the load cell body and the bridge

Shield to body

Resistance between the cable shield (screen) and the load cell body


If the Continuous display mode was selected:

The LED will flash in various colours and the load cell signal will be displayed continuously in
percentage of the full load cell capacity.

For example: applying 10 kg load on a 100 kg load cell will show a value of 10%

Note: To increase the battery lifetime the instrument will shut off after 6 minutes if no key was pressed.