Appendix 5: parameter’s default table, Ppendix, Arameter – Flintec FT-16D V.1.21 User Manual

Page 31: Efault, Able

Advertising
background image

A

PPENDIX

5: P

ARAMETER

S

D

EFAULT

T

ABLE

FT-

Parameter Factory

defaults

16 16D

0-- Interface

x

x

01- Serial Interface 2

x x

10 Data format

0 = Disabled

x

x

11 Baud rate

3 = 9600

x

x

12 Handshake

0 = No

x

x

13 Address

0

x

x

14 Data length and parity

0 = 8 bit, no parity

x

x

02- Serial Interface 3

x x

20 Data format

0 = Disabled

x

x

21 Baud rate

3 = 9600

x

x

22 Handshake

0 = Disabled

x

x

23 Address

0

x

x

24 Mode

0 = RS232C

x

x

25 Data length and parity

0 = 8 bit, no parity

x

x

04- Printer

x x

45 Minimum print

20 kg

x

x

46 Print control

0 = With Print Key

x

x

1-- Configuration

x

x

11- Start

Up

x x

114 Buzzer active?

0 = Disabled

x

x

12- Filter

x x

120 Filter

FT-16 : 7 / FT-16D : 3 x

x

2-- Scale Parameters

x

x

20- Set

up

x x

200 Approved

0 = No

x

x

201 Increased indication

0 = With x10 key

x

x

202 Power on zero

2 = ± %10

x

x

203 Zero with key

1 = ± %2

x

x

204 Auto zero tracking

0 = Disabled

x

x

205 Tare

1 = With Tare Key

x

x

206 Motion detector

1 = ± 0,5e

x

x

21- Scale

Build

x x

210 Scale Type

0 = Single Range

x

x

211 Capacity 1 / d1

x

x

212 Capacity (2) / d(2)

60 t / 20 kg

x

x

22- DLC Configuration Block

x

220 DLC Type

0 = Flintec RC3D

x

221 Quantity of DLC

1

x

222 DLC Address

x

3-- Calibration / Adjustment

x

x

30- Calibration

x x

300 Gravity

x

301 Calibration

x

x

31- Adjustment

x x

310 Zero adjustment

x

x

311 Span adjustment

x

x

312 Span adjustment under load

x

32- Eccentricity

Adjustment

x

320 Adjustment Method

0 = independend DLC

x

321 Auto Adjustment

x

322 Manual Adjustment

x

323 Temporarily set constants to “1” 0 = normal operation

x

8-- Metrological Data

x

x

80- Legal

Metrology

x x

800 Calibration counter

x

x

801 Alibi memory

x

x

802 Calling a record

x

x

803 Print Alibi memory

x

x

804 Alibi memory information

x

x

FT-

Parameter

16 16D

9-- Diagnostics

x

x

90- Tests

x x

900Key Pad test

x

x

904mV indication

x

905DLC internal count

x

91- Serial

numbers

x x

910FT-1x serial no.

x

x

911Option 1 serial no. (if installed)

x

x

912Option 2 serial no. (if installed)

x

x

913Option 3 serial no. (if installed)

x

x

97- Calibration

coefficients

x

97X Calibration coefficients

x

99- Print

Parameters

x x

990All parameters

x

x

991Load default parameters

x

x


FT-16(D) Technical Manual, Rev. 1.21 December 2008