1.

Technical Data

• Weighing range (Max):

150 g

• Readout (d):

0.1 g

• operating conditions:

18 – 25 °C
max. 85% air humidity (not condensing)

• power supply:

2 x 1,5V „AAA“

2.

Operation

• Press the “ON” – key, the balance displays “8.8.8.8” and switches to “0.0 g”.

• Place the weighing object in the middle of the weighing platform; the weight in grams is shown

on the display.

• By pressing “ZERO” the balance can be reset to “0.0 g”.

• Weighing unit switch-over: from g to OZ (button under balance)

3. Package Volume

• Pocket

Balance

• Balance

cover

• Batteries (2 x 1,5V „AAA“)
• Operating

instructions

4. Adjustment

• By pressing “CAL” for 3 seconds until “CAL” appears on the display.

• Let loose the “CAL”-key; the display shows “100”.

• Place recommended adjusting weight (100 g, class M3, not included).

Information concerning the adjusting weights is available at:

http://www.kern-sohn.com

• The display switches to “F” and the adjustment is complete.

• If other messages are on the display, the adjustment was faulty.

Please repeat the adjustment process.

5. Important Notes

• Do not overload the balance (when overloaded, error message “E” appears)

• With the occurrence of electromagnetic fields deviations of the displayed weight value are

possible. In case of deviations change location.

• Avoid high air humidity, vapours and dust, because the balance is not hermetically closed. Do

not bring the balance in direct contact with liquids. Liquids could enter the load cell.

• No permanent load on the weighing platform.

• Never operate the balance in rooms where danger of explosion exists.

• Regularly check the balance with the test weight.