Kern tcb – KERN & SOHN TCB 200-1 User Manual

Page 2

Advertising
background image

GB

KERN TCB

Version 1.2 03/2003

Operating instruction
Pocket balance


1. Technical Data

• Weighing range:

200 g

readability

d:

0.1

g

• operating conditions:

10 – 30 °C
max. 85% air humidity (not condensing)

• power supply:

batteries 2 x LR44

2. Operation

• Turn the cover-disk to position „AUF/OPEN “

• Press the ON/OFF key

; the balance displays “8888” and switches to “0”.

• Place the weighing object in the middle of the weighing platform; the weight in

grams is shown on the display.

• After the weighing process is finished the cover-disk should be turned back to

“ZU/CLOSE”.

• By pressing

the balance can be reset to “0”.


1. Adjustment

• Let loose the ON/OFF-key; the display shows “100”.

• By pressing

for 3 seconds until “CAL” appears on the display.

• Place a 100 g adjusting weight (class M3) on the weighing platform.

• The display switches to “F” and the adjustment is complete.

• If other messages are on the display, the adjustment was faulty. Please repeat

the adjustment process.

2. Important Notes

• Do not overload the balance (when overloaded, error message “E” appears)

• With the occurrence of electromagnetic fields deviations of the displayed

weight value are possible. In case of deviations change location.

• Avoid high air humidity, vapours and dust, because the balance is not

hermetically closed. Do not bring the balance in direct contact with liquids.
Liquids could enter the load cell.

• No permanent load on the weighing platform.

• Never operate the balance in rooms where danger of explosion exists.

• Regularly check the balance with the test weight.

Warranty expires as soon as the balance is opened.

Advertising