AEJ_N/AES_N-BA-e-1321

Enter upper limit „Hi“

HKK

000.0000

g

009.5500

g

Select the digit to be changed by pressing the F-key; each digit currently active
will be flashing.

Select the digit by pressing the UNIT key

Confirm by pressing the PRINT key

From now the balance is in checkweighing mode. Put on goods to be weighed,
tolerance control is started With the help of the indicator check if the weighed goods
are under, inside or over the default tolerance.

Note:
If invalid values are entered such as lower tolerance limit greater than upper
tolerance limit, the balance will issue the error message “Er8 outr“ and return
automatically to weighing mode.


Return to weighing mode