13.6 Pipette calibration

13.6.1 Entering parameters

Prior to measuring you will have to enter the parameters below.

Parameters Description

Set n

Number of measurements per test volume, selectable from 5 -15
measurements

Set t

Test temperature, selectable from 0.4 – 25.5°C

Set P

Air pressure test chamber, selectable from 800 – 1050 hPa

Set h

Air humidity test chamber, selectable from 10 – 95 %

For devices with a fixed volume, test volume equals nominal volume.

For devices with variable volume, the following three volumes need to
be tested:

V

0

The lower limit for the effective volume or 10 % of the nominal volume
[µl], depending on whichever is the larger.

V

1/2

About 50 % of the nominal volume [µl]

V

1

Nominal volume [µl]