13.6.2 Carry out series of measurements for test volume V

0

After entering all parameters (See chap. 13.6.1), position the weighing vessel,
wait for the stability display to appear, then tare by pressing the TARE-key; P1
V

0

will appear, followed by Place 1.

1.

Dispense the test fluid (such as distilled water) into the weighing vessel of
conduit “V

0

”.

2.

Wait for stability display, then press the PRINT button. Place 2 followed by
the current weight display will appear.

3.

Tare, by pressing the TARE-key.

Repeat this test cycle (steps 1 -3) as often as set in “Set n
(see chap. 13.6.1).

After the last test cycle has been confirmed, a display announcing the start of
the second series of measurements “P2 V

1/2

”, followed by Place 1, see chap.

13.6.3, will appear.