KERN & SOHN AEJ 100-4NM User Manual

Page 69

Advertising
background image

AEJ_N/AES_N-BA-e-1321

69

13.6.4 Carry out series of measurements for test volume V

1

Carry out the third series of measurements by following the same sequence of
operations as the previous.

1.

Dispense the test fluid (such as distilled water) into the weighing vessel of
conduit “V

1

” to be measured.

2.

Wait for stability display, then press the PRINT button. Place 2 followed by
the current weight display will appear.

3.

Tare, by pressing the TARE-key.

4.

Repeat this test cycle (steps 1 -3) as often as set in “Set n
(See chap. 13.6.1).

5.

After the last test cycle has been confirmed by pressing the PRINT-key a protocol
will be issued via the RS232 interface. See printout example below.