14 data output rs 232 “p4 print, 1 technical data – KERN & SOHN AEJ 100-4NM User Manual

Page 71

Advertising
background image

AEJ_N/AES_N-BA-e-1321

71

14 Data output RS 232 “P4 Print“

You can print weighing data automatically via the RS 232C interface or manually
pressing the PRINT button via the interface according to the setting in the menu.

This data exchange is asynchronous using ASCII - Code.

The following conditions must be met to provide successful communication between
the weighing system and the printer.

• Use a suitable cable to connect the display unit to the interface of the printer.

Faultless operation requires an adequate KERN interface cable.

• Communication parameters (baud rate, bits and parity) of display unit and

printer must match.

14.1 Technical data

Connection

9 pin d-subminiature bushing

Pin 2: Receive data

Pin 3: Transmit data

Pin 5: Signal ground

Baud rate

2400/4800/9600/19200 selectable

Parity

8 bits, no parity / 1 stop bit

Transfer
modes

• Manually after pressing the PRINT key

• Automatically according to stability display

• Continuously, according to setting

• On request by external device (For remote control commands,

see chap. 14.6)