3 menu settings “p4 print – KERN & SOHN AEJ 100-4NM User Manual

Page 73

Advertising
background image

AEJ_N/AES_N-BA-e-1321

73

14.3 Menu settings “P4 PRINT“

Menu call “P4 PRINT“, see chap. 8.1.

Menu overview

Function

Settings

Description of options

P4.1 bAud

2400 bit/s

Transfer speed

4800 bit/s *
9600 bit/s
19200 bit/s

P4.2 CntA

no*

Continuous output disabled in standard
weighing unit,
Output requires key operation

YES

Continuous output in standard weighing unit

P4.3 Cntb

no*

Continuous output disabled in current weighing
unit, output requires key operation

YES

Continuous output in current weighing unit

P4.4 rEPL

no*

Manual issue after pressing the PRINT key.

YES

Automatic issue of first stable weighing value
Sequence of operations:
1. Taring
2. Place weight, issue of first stable weighing

value

3. Renewed output only possible after weight

was removed. Prerequisite: Display +/- 50
display steps from zero.

4. Place next weight.

P4.5 PStb

Locked in models with
type approval certificate

no

Output even for unstable weighing value

YES*

Output for stable weighing value only

P4.6 Lo

0.01*

Input minimum weight for automatic output:
Weighing value issued automatically if current
weighing value exceeds entered minimum
value. The next weighing value will not be
issued unless the weighing value has
meanwhile dropped below the entered weighing
value.

Factory settings are marked by *.

Advertising
This manual is related to the following products: