2 pc controlled output, 3 output of date/time – KERN & SOHN AEJ 100-4NM User Manual

Page 80

Advertising
background image

80

AEJ_N/AES_N-BA-e-1321

14.5.2 PC controlled output


Response message of balance after remote instruction was sent:
XX_

Instruction:

XX_A CR LF

Instruction accepted; will be executed

XX_ I CR LF

Instruction received; impossible to carry out

XX_^ CR LF

Instruction received but time overflow error occurred

XX_v CR LF

Instruction received, but insufficient load

XX_E CR LF

Error during execution, timeout for stable weighing value
exceeded


Data record format:

1 - 3

4

5

6

7

8 - 16

17

18 - 20

21

22

R

emot

e c

ont

rol

c

omma

nd

B

la

n

k

S

ta

b

ilit

y

d

is

p

la

y

B

la

n

k

S

igns

W

e

ight

v

al

u

e

B

la

n

k

U

n

it

CR

LF



Instruction:

1. up to 3 signs

Stability display

Blank character for stable value

? For instable value

^ for excess load

v for underload

Signs

Blank character for positive value

Minus sign for negative value

Weight value

9 signs, right justified

Unit

3 signs, left justified

14.5.3 Output of date/time


The display of date and time can be enabled under menu item “P2 GLP
(see chap. 12):

PdAt – yes

Ptin - yes


Advertising