KERN & SOHN DS User Manual

Page 21

Advertising
background image

KERN

FKB

8K0.1M

FKB

65K1M

Accuracy class

II

II

Readability (d)

0.1 g

1 g

Verification value (e)

1.0 g

10 g

Weighing range (max)

8 000 g

65 000 g

Minimum load (Min)

5.0 g

50 g

Taring range (subtractive)

8 000 g

65 000 g

Reproducibility

0.1 g

1 g

Linearity

± 0.3 g

± 3 g

Smallest piece weight

0.1 g

1 g

Adjustment points

2/5/7/8 kg

20/30/50/60 kg

Recommended adjusting
weight F1 (not supplied)

5 kg + 2 kg

50 kg

Humidity of air

max. 80% rel. (non-condensing)

Stabilization time (typical)

3 sec.

Permitted environmental
temperature

+10 °C ... + 30 °C

+10 °C ... + 30 °C

Warm-up time

2 hours

2 hours

Housing (B x D x H) mm

350 x 390 x 120

Vibration filter

yes

Weighing plate stainless
steel mm

340 x 240

340 x 240

Units

see menu

see menu

Weight kg (net)

6.5

6.5

Data interface

yes (RS232)

Battery operation

6 x 1,5 V; Size C


572/573/KB/DS/FKB/KBJ-BA-e-1472