12 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.


Help:

Fault

Possible cause

The displayed weight does not glow.

• The balance is not switched on.
• The mains supply connection has been

interrupted (mains cable not plugged
in/faulty).

• Power supply interrupted.

Batteries are inserted incorrectly or empty

• No batteries inserted.

The displayed weight is permanently
changing

• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with

foreign matter.

• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)

The weighing value is obviously

wrong

• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging

(choose different location/switch off
interfering device if possible)


Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.