KERN & SOHN FKB 65K1A User Manual

Page 37

Advertising
background image

rE Cr:

The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.

Take into account that the following remote control commands must be sent
without a subsequent CR LF.


s

Function:

Stable weighing value for the weight is sent via the RS232
interface

w

Function:

Weighing value for the weight (stable or unstable) is sent
via the RS232 interface

t

Function:

No data are sent, the balance carries out the tare function.


h. Format for stable values for weight/quantity/percentage
1

2

3

4

5

6

7

8

9

10 11 12 13 14 15 16 17 18

B

*

B

B

B

B

B

B

B

B

0

.

0

B

g

B

B

CR LF


i. Format in case of fault
1

2

3

4

5

6

7

8

9

10

11

12

13

14

B

B

B

B

B

B

B

E

r

r

o

r

CR LF


j. Format for unstable values for weight/quantity/percentage
1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

B

*

B

B

B

B

B

B

B

B

0

.

0

B

CR LF



Symbols


B

*

Blank or M

B / 0 /. Blank / weight values / decimal dot, depending on weighing value

g

Weight unit / pcs / %

E, o, r

ASCII code or “E, o, r”

CR

Carriage return symbol

LF

Line feed symbol